TemperaturprøvelokaleCTS
TSR-30/1000
Temperaturprøvelokale
CTS
TSR-30/1000
Stand
Brugt
Beliggenhed
Borken 

Vis billeder
Vis kort
Oplysninger om maskinen
Pris og placering
- Beliggenhed:
- Einsteinstraße 8a, 46325 Borken, Deutschland

Ring op
Detaljer om tilbuddet
- Annonce-ID:
- A14926978
- Referencenummer:
- 23988
- Opdatering:
- sidst den 15.11.2024
Beskrivelse
CTS
Type: TSR-30/1000
Water-cooled
The type designation of the CTS cabinet series is structured as follows:
Example: Type TSR-30/1000
1) T stands for: Temperature
2) S stands for: Stress Screening
3) R stands for: External Conditioning
4) -30 indicates: lower temperature limit -30 °C
5) 1000 indicates: 1000 liters test chamber volume
Technical Performance Data:
A) TSR-30/1000 base unit without hose connections,
i.e., when used as a standalone temperature test chamber
Temperature testing:
Temperature range: -40°C to +180°C
Final temperature: -45°C
Temperature deviation: ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5
Heating: 6.0 K/min
Cooling: 5.0 K/min
Heat radiation:
Approx. 5000 W at +20°C
Approx. 4200 W at ±0°C
Approx. 3000 W at -20°C
Technical data are based on ambient temperature of +25°C,
Nominal voltage 400V, without test specimen, without heat input and without additional equipment.
Air volume flow
Approx. 2000 m³/h
B) TSR-30/1000 base unit with hose connections,
for connection to a customer-provided external test chamber
Temperature testing:
Temperature range: -30°C to +170°C
Temperature deviation: ≤ ±0.1 to 0.5 K over time
Average temperature change rate,
Itsdpfx Absrivfcsysd
measured with a closed-loop circuit
Heating: 4.0 K/min
Cooling: 4.0 K/min
within the range +150°C to -10°C
Technical data are based on ambient temperature of +25°C,
Nominal voltage 400V, without test specimen, without heat input and without additional equipment,
measured in a closed-loop circuit
Note: The customer-provided external test chamber must be completely sealed and properly insulated;
any penetrations must be optimally sealed by the customer.
Dimensions:
Test volume: 1000 liters
Recirculating air connection: Ø 200 mm
Test space dimensions: Height 900 mm, Width 1000 mm, Depth 1100 mm
External dimensions: Height 1950 mm, Width 1270 mm, Depth 2245 mm
Technical Design:
Refrigeration unit:
water-cooled, low-noise full hermetic design
stepless performance adjustment via
electronic monitoring and control system,
environmentally friendly refrigerant R404A
Door:
fully openable, left-hinged door,
one-hand operation, lockable.
Shelf grate:
stainless steel, support area 970 x 980 mm
Load capacity per shelf max. 35 kg
Total load with max. shelves: 100 kg
Test chamber lighting: halogen lamp 20 W, 12 V
Control system:
- 32-bit controller
- LCD control panel
Operating data:
Electrical connection: 400 V +6/-10 %, 3/N, 50 Hz
Rated power: 13.3 kW
Protection class: IP 22
Sound pressure level: < 60 dB(A)
Weight: approx. 850 kg
For your safety as a buyer, please note the following information!
The following points are checked or serviced in advance on our offered chambers:
1. Function check and replacement of necessary components
2. If required, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers undergo a documented test run.
Condition: used
Scope of delivery: (see image)
(Alterations and errors in the technical data and information reserved!)
For further questions, we are happy to assist you by phone.
Annoncen er blevet oversat automatisk. Oversættelsesfejl kan forekomme.
Type: TSR-30/1000
Water-cooled
The type designation of the CTS cabinet series is structured as follows:
Example: Type TSR-30/1000
1) T stands for: Temperature
2) S stands for: Stress Screening
3) R stands for: External Conditioning
4) -30 indicates: lower temperature limit -30 °C
5) 1000 indicates: 1000 liters test chamber volume
Technical Performance Data:
A) TSR-30/1000 base unit without hose connections,
i.e., when used as a standalone temperature test chamber
Temperature testing:
Temperature range: -40°C to +180°C
Final temperature: -45°C
Temperature deviation: ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5
Heating: 6.0 K/min
Cooling: 5.0 K/min
Heat radiation:
Approx. 5000 W at +20°C
Approx. 4200 W at ±0°C
Approx. 3000 W at -20°C
Technical data are based on ambient temperature of +25°C,
Nominal voltage 400V, without test specimen, without heat input and without additional equipment.
Air volume flow
Approx. 2000 m³/h
B) TSR-30/1000 base unit with hose connections,
for connection to a customer-provided external test chamber
Temperature testing:
Temperature range: -30°C to +170°C
Temperature deviation: ≤ ±0.1 to 0.5 K over time
Average temperature change rate,
Itsdpfx Absrivfcsysd
measured with a closed-loop circuit
Heating: 4.0 K/min
Cooling: 4.0 K/min
within the range +150°C to -10°C
Technical data are based on ambient temperature of +25°C,
Nominal voltage 400V, without test specimen, without heat input and without additional equipment,
measured in a closed-loop circuit
Note: The customer-provided external test chamber must be completely sealed and properly insulated;
any penetrations must be optimally sealed by the customer.
Dimensions:
Test volume: 1000 liters
Recirculating air connection: Ø 200 mm
Test space dimensions: Height 900 mm, Width 1000 mm, Depth 1100 mm
External dimensions: Height 1950 mm, Width 1270 mm, Depth 2245 mm
Technical Design:
Refrigeration unit:
water-cooled, low-noise full hermetic design
stepless performance adjustment via
electronic monitoring and control system,
environmentally friendly refrigerant R404A
Door:
fully openable, left-hinged door,
one-hand operation, lockable.
Shelf grate:
stainless steel, support area 970 x 980 mm
Load capacity per shelf max. 35 kg
Total load with max. shelves: 100 kg
Test chamber lighting: halogen lamp 20 W, 12 V
Control system:
- 32-bit controller
- LCD control panel
Operating data:
Electrical connection: 400 V +6/-10 %, 3/N, 50 Hz
Rated power: 13.3 kW
Protection class: IP 22
Sound pressure level: < 60 dB(A)
Weight: approx. 850 kg
For your safety as a buyer, please note the following information!
The following points are checked or serviced in advance on our offered chambers:
1. Function check and replacement of necessary components
2. If required, new filling with legally compliant refrigerant
3. Leak test with certificate
4. After successful inspection, the chambers undergo a documented test run.
Condition: used
Scope of delivery: (see image)
(Alterations and errors in the technical data and information reserved!)
For further questions, we are happy to assist you by phone.
Annoncen er blevet oversat automatisk. Oversættelsesfejl kan forekomme.
Send anmodning
Telefon & Fax
+49 2861 ... annoncer
Du kan også være interesseret i disse annoncer.
Lille annonce
Emmelshausen
693 km
Drejebænke
GildemeisterCTX beta 1250 4 A
GildemeisterCTX beta 1250 4 A
Lille annonce
Berlin
490 km
Vertikal sprøjtestøbemaskine
ENGELe-insert 740H/60 single-C
ENGELe-insert 740H/60 single-C
Lille annonce
Deutschland
571 km
Sprøjtestøbemaskine
ArburgALLROUNDER CENTEX 320C 500-250 VO
ArburgALLROUNDER CENTEX 320C 500-250 VO
Lille annonce
Beringen
956 km
Pallebærer
MIKRONHSM 400
MIKRONHSM 400
Lille annonce
Meiningen
637 km
Lakttørrer DIN 1539 300°C
VötschVTL 100/150 nach DIN 1539
VötschVTL 100/150 nach DIN 1539
Lille annonce
Pfullingen
869 km
Bearbejdningscenter
ABENEVHF 330 TNC 620
ABENEVHF 330 TNC 620
Lille annonce
Hessisch Lichtenau
564 km
CNC fræsemaskine med HEIDENHAIN TNC 320, universal fræsemaskine, konsolfræsemaskine.
ABENEVHF-360TI
ABENEVHF-360TI
Lille annonce
Sankt Stefan
1.115 km
Bearbejdningscenter
MikronHSM 800
MikronHSM 800
Lille annonce
Geertruidenberg
592 km
Klimakammer
Feutron3703/13
Feutron3703/13
Lille annonce
Borken
520 km
Klimaprøvelokale
VötschVCS 7033
VötschVCS 7033
Din annonce er blevet slettet